FP640 FP6430 Touch Screen Flame Photometer Price

7 inch touch screen
Direct concentration display(excluding FP640)
Auto calculate correlation coefficient(excluding FP640)
Store 200 measuring data and 10 standard curve(excluding FP640/FP6410)

Product Description:

What is Flame photometer?

Flame photometer is an instrument for quantitative analysis of alkali metal and alkaline earth metal elements in samples according to the characteristic wavelength spectrum of the atom or ion of the element to be measured after being excited by flame.

The flame photometer itself cannot obtain the absolute concentration value of the measured element. You must first prepare a standard solution, calibrate photometer, draw a standard curve, and then measure the unknown solution. After obtaining the reading displayed by the instrument, find the corresponding concentration value from the curve, so as to obtain the unknown concentration value of the measured substance.

Flame photometer Application?

Flame photometer has the characteristics of high sensitivity, strong selectivity, less samples required and fast analysis speed, and has been widely used in medical and clinical, soil, fertilizer, glass, ceramics, cement, refractory materials, and silicic acid industry.

Features:

  • 7 inch touch screen

  • Direct concentration display(excluding FP640)

  • Auto calculate correlation coefficient(excluding FP640)

  • Store 200 measuring data and 10 standard curve(excluding FP640/FP6410)

  • Pre-selection of flame sizes

  • Flame out protection device

  • Measuring range changing

  • Concentration units selectable

  • Multilingual user interface

Specifications:

ModelFP640FP6410FP6430FP6431FP6440FP6450
Display7 inch touch screen




Dislay dataOptical powerConcentration valueConcentration valueConcentration valueConcentration valueConcentration value
Test elementK, NaK, NaK, Na, LiK, Na, Ca

K, Na, Li,

Ca

K, Na, Li,

Ca, Ba

Range

ppm

K0-1000-1000-1000-1000-1000-100

Na0-1600-1600-1600-1600-1600-160

Li//0-100/0-1000-100

Ca///0-10000-10000-1000

Ba/////0-3000

LOD

ppm

K0.010.010.010.010.010.01

Na0.010.010.010.010.010.01

Li//0.1/0.10.1

Ca///222

Ba/////6

Linear

error

K0.1950.1950.1950.1950.1950.195

Na0.690.690.690.690.690.69

Li//0.15/0.150.15

Ca///333

Ba/////9
Channel Qty.223345
Response time<8s<8s<8s<8s<8s<8s
Sample uptake<6ml/min<6ml/min<6ml/min<6ml/min<6ml/min<6ml/min
Stability< 3% drift over15s when continuously aspirating.
Reproducibility< 3% coefficient of variation for 7 consecutive samples
Curve Graph//



build-in thermal printer/OptionalOptionalOptionalOptionalOptional
Output/USBUSBUSBUSBUSB
FuelLPGLPGLPGLPGLPGLPG
PowerAC220V±22V 50Hz±1Hz,250W
Shipping size570*530*400mm/18kg